Testomat 2000® CLF
Artikelnummer: 100240
* Preise exkl. MwSt. zzgl. Versandkosten.

Produktinformationen
Maße: 380 x 280 x 480 mm
Gewicht: 10,600 kg
Messparameter: Free chlorine
Messbereich: 0.00 – 0.99 mg/l (Resolution: 0.01 mg/l) 1 – 2.5 mg/l (Resolution: 0.1 mg/l)
Einsatzbereich: Chlorination monitoring for drinking-water or swimming pools. | Provides protection for reverse osmosis membranes. | Control of chlorine biocides and conditioning agents.
Menüführung: German, English, French
Beschreibung
Beschreibung "Testomat 2000® CLF"
The Testomat 2000® CLF process photometer is an online analyzer for monitoring the ontent of free chlorine (CLF) in the range of 0-2.5 mg / L (ppm).
Performance profile:
- Analysis via automatic dosing of two reagents
- The analysis result is displayed after a reaction time of approx. 1 min. (Measuring time without rinsing time)
- 2 independently programmable limit contacts for monitoring and control tasks
- Menu-driven operation and programming via plain text display
- High-precision titration using a piston dosing pump
- Analysis initiation:
- Automatic interval operation
- (Interval pause adjustable from 0–99 min)
- External control
- Dynamic (fatigue-dependent interval operation)
- Quantity dependent via contact water meter
- Two independent limits with hysteresis (1, 2 oder 3 bad analyzes) and adjustable switching functions
- Monitoring of two measuring points (Switching by external solenoid valves)
- programmable service address
- Internal error documentation
- Programmable maintenance interval for maintenance request