Performance/execution: 115 V/50-60 HZ, 230 V/50-60 HZ, 24 V/50-60 HZ
Language in the display: Dutch, English, French, German, Spanish

Product information

Dimensions: 380 x 280 x 480 mm
Weight: 10,600 kg
Measurement parameters: Phosphate PO4
Measuring range: 0.0 – 10.0 mg/l Resolution: 0.1 mg/l in the range of 0.0 – 7 mg/l 0.25 mg/l in the range of 7 – 10 mg/l Limit of detection: 0,1 or 0,25 mg/l depending on range Accuracy: < 10% at end of measuring range
Range of application: treated wastewater (sewage treatment plants) | online – environmental analysis | monitoring of conditioning agents in cooling and heating circuits
Menu navigation: German, English, French, Dutch, Spanish

Description

Description "Testomat 2000® PO4"

The Testomat 2000® PO4 from Heyl reliably and continuously monitors the orthophosphate content in the range from 0 to 10.0 mg/l (ppm). The analysis method is based on the molybdate blue method. This involves a chemical reaction that produces a characteristic steel blue color change in the sample. The color intensity is evaluated photometrically. The Testomat thus automatically delivers accurate, stable measurements – ideal for the continuous monitoring of process and service water, e.g., in the feed water of a steam boiler or the precipitation tank of a sewage treatment plant.

Performance profile:

  • Analysis by automatic addition of 2 reagents
  • Measurement result display after a reaction time of approx. 10 minutes (measurement time without rinsing time)
  • Menu-guided operation and programming via plain text display
  • Selection of units in ppm or mg/l
  • Analysis trigger:

o Automatic interval operation (interval pause adjustable from 0-255 minutes)

o External control

o Quantity-dependent via contact water meter

  • Two independent limit values with hysteresis (1, 2, or 3 poor analyses) and adjustable switching functions
  • Monitoring of two measuring points (switching via external solenoid valves)
  • Internal error documentation
  • Programmable service address
  • Programmable maintenance interval for maintenance requests